Health product recall

Laerdal Compact Suction Unit 4 (LCSU 4)

Starting date:
December 7, 2016
Posting date:
January 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61636

Reason

The on/off functionality in the LCSU 4 is controlled by electronic components. One of these components might be damaged. The damage occurs over time when the unit is connected to a power source (battery or external power source), but is not related to frequency of use.

A damaged component will result in failure of the on/off functionality: the LCSU 4 cannot be turned off, and it may turn on by itself.

Affected products

Laerdal Compact Suction Unit 4 (LCSU 4)

Lot or serial number

150505A0836 to 160901A0226

Model or catalog number

88005101

88005201

88006101

88006201

Companies

Manufacturer

Laerdal Medical AS

Tanke Svilandsgate 30, P.O. Box 377,

Stavanger, Rogaland Fylke

4002

NORWAY