This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

LABType HD Class II DRB1 Typing Test

Starting date:
March 30, 2012
Posting date:
November 26, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15886

Recalled Products

A. LABType HD Class II DRB1 Typing Test

Reason

The user documenation for this product is missing DRB1*15:50N Allele for Probe on Bead #525.  This may generate a sample typing of DRB1*15:04 instead of DRB1*15:04/15:50N.  There is low risk of death or serious harm to the patient or end user as a result of this potential product failure.

Affected products

A. LABType HD Class II DRB1 Typing Test

Lot or serial number

07B

Model or catalog number
  • RSSOH2B1
  • RSSOH2B1X
Companies
Manufacturer
One Lambda Inc.