Klebsiella Pneumoniae Derived from NCTC 13439 (2020-10-19)
- Starting date:
- October 19, 2020
- Posting date:
- October 30, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74215
Last updated:
Affected Products
Klebsiella Pneumoniae Derived from NCTC 13439
Reason
A complaint investigation identified that the lot numbers listed above do not contain the VIM-1 gene and are not metallo beta-lactamase positive. Identification of the loss of the VIM-1 gene was confirmed via Verigene Gram-Negative Blood Culture Nucleic Acid Test, metallo beta-lactamase (MBL) IP/IPI Etest and mCIM/eCIM testing. If this product is not used specifically for VIM-1/metallo beta-lactamase detection there will be no impact on the results.
Affected products
Klebsiella Pneumoniae Derived from NCTC 13439
Lot or serial number
1245-03-1
1245-03-2
1245-03-3
1245-03-4
1245-04-1
Model or catalog number
01245P
Companies
- Manufacturer
-
Microbiologics, Inc.
200 Cooper Ave. N
St. Cloud
56303
Minnesota
UNITED STATES