Product
Karl Storz Injection Needles
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Flexible Injection Needle

All lots.

27182A

Injection Needle

All lots.

26175 PD
810506
8598D

Laryngeal Injection Needle

All lots.

8598B
8598A

Issue

Validation testing of the efficacy of the manual cleaning for the injection needle was performed. The testing showed that the required efficacy level was not achieved.  Different scenarios were tested, and all possible manual cleaning steps included, and the residual protein value remained above the acceptance criteria.

Recall start date: October 16, 2023