Health product recall

ITS4850A IntelliVue Telemetry System - Monitor (2019-03-25)

Starting date:
March 25, 2019
Posting date:
May 3, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69810

Last updated: 2019-05-03

Affected Products

ITS4850A IntelliVue Telemetry System - Monitor

Reason

Problem 1
some MX4O patient worn monitors may have excessive battery power consumption. For those MX4O's with revision B software and fully charged batteries may last approximately 25% shorter than is specified in the MX4O instructions for use. This issue affects units operating on both rechargeable and disposable batteries.
Problem 2
Low battery technical alerts may not occur as specified in the instructions for use.
For those MX4O's that:
1) operate with disposable AA batteries, and
2) include optional SP02 used in automatic measurement mode.
The technical alerts that may be absent include:
- "TELE BATTERY LOW",
- "NO SPO2T: BATT LOW",
- "REPLACE TELE BATTERY"

Affected products

ITS4850A IntelliVue Telemetry System - Monitor

Lot or serial number

Not applicable.

Model or catalog number

MX40

Companies
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES