ITS4850A IntelliVue Telemetry System - Monitor (2019-03-25)
- Starting date:
- March 25, 2019
- Posting date:
- May 3, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69810
Last updated: 2019-05-03
Affected Products
ITS4850A IntelliVue Telemetry System - Monitor
Reason
Problem 1
some MX4O patient worn monitors may have excessive battery power consumption. For those MX4O's with revision B software and fully charged batteries may last approximately 25% shorter than is specified in the MX4O instructions for use. This issue affects units operating on both rechargeable and disposable batteries.
Problem 2
Low battery technical alerts may not occur as specified in the instructions for use.
For those MX4O's that:
1) operate with disposable AA batteries, and
2) include optional SP02 used in automatic measurement mode.
The technical alerts that may be absent include:
- "TELE BATTERY LOW",
- "NO SPO2T: BATT LOW",
- "REPLACE TELE BATTERY"
Affected products
ITS4850A IntelliVue Telemetry System - Monitor
Lot or serial number
Not applicable.
Model or catalog number
MX40
Companies
- Manufacturer
-
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES