ITS4850A Intellivue Telemetry System (2020-02-28)
- Starting date:
- February 28, 2020
- Posting date:
- March 13, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72589
Last updated: 2020-03-13
Affected Products
ITS4850A Intellivue Telemetry System
Reason
The ECG signal from patients being monitored using a Philips TRx4851A Telemetry Transceiver may not be properly analyzed when the transceiver is used with a Philips Patient Information Center iX Release C.02.00, C.02.02, C.02.03 (all released versions of C.02, collectively referred to as âC.02.xx’) or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
Alarming based on pulse oximetry (SpO2) signals are unaffected by this issue.
Affected products
ITS4850A Intellivue Telemetry System
Lot or serial number
Not applicable.
Model or catalog number
TRX4851A
Companies
- Manufacturer
-
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES