This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Irrigation Set

Starting date:
July 1, 2016
Posting date:
March 16, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62654

Reason

Omnia has determined that during the assembling phase, if tightening rings are not perfectly placed, it is possible that in a condition of maximum flow and pressure the tube can detach causing leakage of saline solution.

Affected products

Irrigation Set

Lot or serial number

15040308

15042402

15111204

Model or catalog number

32.F0134.00

Companies

Manufacturer

Omnia Spa

Via Francesco Delnevo 190sx

Fidenza PR

43036

ITALY