This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

iQ200 Sprint, Select, and Elite Urine Microscopy Analyzers (Pro and Plus) (2017-11-13)

Starting date:
November 13, 2017
Posting date:
December 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-65456

Affected products

  1. IQ 200 SPRINT AUTOMATED URINE MICROSCOPY ANALYZER
  2. IQ200 SELECT URINALYSIS SYSTEM - ANALYZER
  3. IQ 200 ELITE AUTOMATED URINE MICROSCOPY ANALYZER
  4. IQ 200 ELITE PLUS
  5. IQ 200 ELITE PRO
  6. IQ 200 SELECT PRO
  7. IQ 200 SELECT PLUS
  8. IQ 200 SPRINT PRO
  9. IQ 200 SPRINT PLUS

Reason

Iris International has determined that the iQ200 Series Urine Microscopy Analyzer with barcode reader (model NFT-2100), may intermittently exhibit the following:

  • Failure to read the urine sample dilution barcode label where it defaults to a dilution factor of 1:1 and does not apply the correct dilution factor.
  • Failure to read body fluid dilution barcode label where the rack is rejected and no results are generated.

Affected products

A. IQ 200 SPRINT AUTOMATED URINE MICROSCOPY ANALYZER

 

Lot or serial number

Not Applicable

Model or catalog number

700-3322

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



B. IQ200 SELECT URINALYSIS SYSTEM - ANALYZER

 

Lot or serial number

7345

7369

8581

8682

8969

9624

Model or catalog number

800-3900

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



C. IQ 200 ELITE AUTOMATED URINE MICROSCOPY ANALYZER

 

Lot or serial number

Not Applicable

Model or catalog number

800-3920

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



D. IQ 200 ELITE PLUS

 

Lot or serial number

3999

6996

7305

7831

9817

Model or catalog number

800-3934

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



E. IQ 200 ELITE PRO

 

Lot or serial number

3999

6996

7305

7831

9817

Model or catalog number

800-3052

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



F. IQ 200 SELECT PRO

 

Lot or serial number

Not Applicable

Model or catalog number

800-3051

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



G. IQ 200 SELECT PLUS

 

Lot or serial number

5883

5938

5985

6215

8057

8089

8324

Model or catalog number

800-3935

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



H. IQ 200 SPRINT PRO

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

800-3053

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



I. IQ 200 SPRINT PLUS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

800-3933

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES