IPRO CARELINK THERAPY MANAGEMENT SOFTWARE (2017-10-10)
- Starting date:
- October 10, 2017
- Posting date:
- November 10, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-65062
Affected products
IPRO CARELINK THERAPY MANAGEMENT SOFTWARE
Reason
Due to a time conversion error, data uploaded from the iPro2 recorder to the CareLink iPro software using the new CareLink iPro uploader feature may result in incorrect reports. This pertains only to reports generated between September 23, 2017 and September 27, 2017 that used the new CareLink iPro uploader feature. Specifically, when data from the iPro2 recorder was uploaded using the new CareLink iPro uploader feature, the glucose reading timestamps inadvertently defaulted to Greenwich Mean Time (GMT) /Coordinated Universal Time (UTC). Manufacturer determined that the difference in time zones could result in calculation and display errors on the reports. This could lead to potential differences or inaccuracies in the sensor tracings, which also affect the display of meals and activities and the resulting report.
Affected products
IPRO CARELINK THERAPY MANAGEMENT SOFTWARE
Lot or serial number
3030864
3058602
3075370
3081022
3087942
3088123
3088217
Model or catalog number
MMT-7340
Companies
- Manufacturer
-
MEDTRONIC MINIMED