This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

IPRO CARELINK THERAPY MANAGEMENT SOFTWARE (2017-10-10)

Starting date:
October 10, 2017
Posting date:
November 10, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-65062

Affected products

IPRO CARELINK THERAPY MANAGEMENT SOFTWARE

Reason

Due to a time conversion error, data uploaded from the iPro2 recorder to the CareLink iPro software using the new CareLink iPro uploader feature may result in incorrect reports. This pertains only to reports generated between September 23, 2017 and September 27, 2017 that used the new CareLink iPro uploader feature. Specifically, when data from the iPro2 recorder was uploaded using the new CareLink iPro uploader feature, the glucose reading timestamps inadvertently defaulted to Greenwich Mean Time (GMT) /Coordinated Universal Time (UTC). Manufacturer determined that the difference in time zones could result in calculation and display errors on the reports. This could lead to potential differences or inaccuracies in the sensor tracings, which also affect the display of meals and activities and the resulting report.

Affected products

IPRO CARELINK THERAPY MANAGEMENT SOFTWARE

Lot or serial number

3030864

3058602

3075370

3081022

3087942

3088123

3088217

Model or catalog number

MMT-7340

Companies
Manufacturer

MEDTRONIC MINIMED

18000 DEVONSHIRE ST

NORTHRIDGE

91325

California

UNITED STATES