IOL Master 700 (2015-02-04)
- Starting date:
- February 4, 2015
- Posting date:
- July 22, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-59502
Affected products
A. IOL Master 700
Reason
The current IOLmaster 700 software version 1.14.84 sometimes shows an unexpected behavior during toric IOL power calculation with the haigis suite: in some cases of low cylinder power with a theoretically ideal IOL below the delivery range of the selected toric IOL model, the device suggests a toric IOL with significant remaining cylinder refraction.
Affected products
A. IOL Master 700
Lot or serial number
1128467
1128480
1131493
1131495
1131507
1131532
1131535
1131539
Model or catalog number
IOL MASTER 700
Companies
- Manufacturer
-
Carl Zeiss Meditech AG
Goeschwitzer Strasse 51-52
Jena
07745
GERMANY