Invia Motion Negative Pressure Wound Therapy (NPWT) System (2018-06-05)
- Starting date:
- June 5, 2018
- Posting date:
- July 5, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67170
Affected products
Invia Motion Negative Pressure Wound Therapy (NPWT) System
Reason
Medela AG identified an issue where the unforeseen user battery charging behaviour has led to a malfunction. This malfunction was identified only in the case where the user repeatedly plugged and unplugged the pump to the power source and the battery was fully charged. Technically, this can cause the safety fuse to temporarily open the battery connection. When this situation occurs, the device no longer recognises the presence of the battery and notifies the user with a visual and audible "battery missing" notification. In case the device is unplugged in this situation, this will cease operation of the device.
Affected products
Invia Motion Negative Pressure Wound Therapy (NPWT) System
Lot or serial number
1676710
Model or catalog number
087.4012