Health product recall

Invia Motion Negative Pressure Wound Therapy (NPWT) System (2018-06-05)

Starting date:
June 5, 2018
Posting date:
July 5, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67170

Affected products

Invia Motion Negative Pressure Wound Therapy (NPWT) System

Reason

Medela AG identified an issue where the unforeseen user battery charging behaviour has led to a malfunction. This malfunction was identified only in the case where the user repeatedly plugged and unplugged the pump to the power source and the battery was fully charged. Technically, this can cause the safety fuse to temporarily open the battery connection. When this situation occurs, the device no longer recognises the presence of the battery and notifies the user with a visual and audible "battery missing" notification. In case the device is unplugged in this situation, this will cease operation of the device.

Affected products

Invia Motion Negative Pressure Wound Therapy (NPWT) System

Lot or serial number

1676710

Model or catalog number

087.4012

Companies
Manufacturer
Medela AG
Lättichstrasse 4B
BAAR, ZUG
6341
SWITZERLAND