This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

INVACARE PRECISE RX PEDIATRIC FLOWMETER (2016-06-30)

Starting date:
June 30, 2016
Posting date:
July 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59460

Affected products

A. INVACARE PRECISE RX PEDIATRIC FLOWMETER

Reason

The rear gradation scale was screen printed upside down on the flowmeter gauge. Proper delivery of oxygen is critical to infants. In the population at greatest risk i.e. pediatric population, health impact specific to high oxygen saturation could result in systemic toxicity, potential lung vascular injury, or over extended period of time could cause blindness.

Affected products

A. INVACARE PRECISE RX PEDIATRIC FLOWMETER

Lot or serial number

16BF017557

16CF030112

16CF030121

16CF032024

16CF040071

16CF040083

16DF040094

 

Model or catalog number

IRCPF16

IRCPF16AW

Companies
Manufacturer

INVACARE CORPORATION

2101 E. LAKE MARY BLVD.

SANFORD

32773

Florida

UNITED STATES