This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Invacare Perfect O2 and Invacare Platinum Oxygen Concentrator (2014-12-29)

Starting date:
December 29, 2014
Posting date:
January 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43155

Recalled Products

A) Invacare Perfect O2

B) Invacare Platinum Oxygen Concentrator

Reason

Invacare has learned that the psi (pressure sensitive interrupter) designed for the compressor motor start/run capacitor by one supplier could lead to rupture. That capacitor in the oxygen concentrator has an internal pressure relief component (the psi) which may not operate as intended permitting the capacitor to rupture in certain circumstances. If the psi does not function to relieve the heat and pressure due to a short circuit within the capacitor, then the pressure can cause a rupture in the seam between the housing and the end plate. If the rupture occurs, the oil may be expelled from the capacitor which may result in a fire hazard.

Affected products

A) Invacare Perfect O2

Lot or serial number

10CSZ43XXXX - 12GSZ62XXXX

10FSZ470209

10HSZ470004

10ISZ470005

10ISZ470051

Model or catalog number

IRC5P

IRC5PO2

IRC9LXO2AWQN

Companies
Manufacturer

Invacare Corp.

One Invacare Way

Elyria

44036

Ohio

UNITED STATES



B) Invacare Platinum Oxygen Concentrator

Lot or serial number

10CSZ43XXXX - 12GSZ62XXXX

10FSZ470209

10HSZ470004

10ISZ470005

10ISZ470051

Model or catalog number

IRC5P

IRC5PO2

IRC9LXO2AWQN

Companies
Manufacturer

Invacare Corp.

One Invacare Way

Elyria

44036

Ohio

UNITED STATES