InTouch Software Repair Kit (2019-10-29)
- Starting date:
- October 29, 2019
- Posting date:
- November 22, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71670
Last updated: 2019-11-22
Affected Products
InTouch Software Repair Kit
Reason
Via customer complaints and subsequent non-conformance review [March 2019], it has been identified that the audible bed exit alarm is affected by a software issue in the footboard that prevents users from being able to change the alarm volume. The impacted footboard software is REV_0665-C. If a user attempts to adjust the bed exit alarm volume, the slider may indicate that it is set to high volume, however, the audible alarm will sound at the low volume setting. The volume then cannot be changed by the user
Affected products
InTouch Software Repair Kit
Lot or serial number
All lots.
Model or catalog number
27-2893K
Companies
- Manufacturer
-
Stryker Medical