Health product recall

Intellanav Stablepoint Ablation Catheter

Last updated

Summary

Product
Intellanav Stablepoint Ablation Catheter
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Intellanav Stablepoint Ablation Catheter Standard Curve

More than 10 numbers, contact manufacturer.

M004ERFSDS96200

Intellanav Stablepoint Ablation Catheter Large Curve

More than 10 numbers, contact manufacturer.

M004ERFSDS9620K20

Issue

During the execution of 2 year real time aging, an issue was identified during bubble testing where there was a leak in the sealed sterile barrier of the packaging. Further investigation determined the manufacturing work instruction was not aligned to the design drawing. It was determined during review of the dv failure that the breach of the sterile barrier was related to the incorrect placement of the IFU during the packaging process.

Recall start date: March 24, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Boston Scientific Corporation

300 Boston Scientific Way, Marlborough, Massachusetts, United States, 01752

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73464

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe