Intelepacs
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
---|---|---|
Intelepacs |
Version 4-10-1 and above |
INTELEPACS |
Issue
There is a possibility that a patient's final report sent out of Intelepacs contains the preliminary text instead of the final report text. This issue can manifest itself only:
- In specific pacs-driven workflow configurations (e.g. multi-method reporting with transcriptionist worklist) where the signing step is handled by Intelepacs,
- Specifically when a network connectivity problem occurs between the server the user is connected to and the primary RIS gateway server (rig) of the system, and
- When the rig servers are not fully synchronized.
Recall start date: Apr 25, 2022