This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Instrument Manager Software (Using Roche Cobas E411 Driver)

Starting date:
December 3, 2012
Posting date:
February 4, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-20221

Recalled Products

A. Instrument Manager Software (Using Roche Cobas E411 Driver)

Reason

When using the Roche Cobas E411 (RCHCE41I) driver version 8.00.0001 or 8.00.0002, if there is a barcode read error, the current sample may be identified as the last sample ID that was in that rack or disk position potentially causing sample mismatch. If the last sample had any pending tests in the orders database, test selection would be downloaded to the instrument with the incorrect sample ID, and results would be reported on the wrong sample.

Affected products

A. Instrument Manager Software (Using Roche Cobas E411 Driver)

Lot or serial number
  • Version 8.00.0001
  • Version 8.00.0002
Model or catalog number
  • 8.00.0001
  • 8.00.0002
Companies
Manufacturer

Data Innovations LLC