Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-25911

Recalled Products

  1. Instatrak 3500 Plus System

Reason

The reason the recall was intiated was because several potential safety issues have been identified with the use of the device. The navigation system may lock up or restart unexpectedly. If two scans from the same modality and same patient are transferred to the system, they may combine into one scan, and the transmitter arm movement after registration may result in inaccuracy and users may experience difficulty calibrating the device.

Affected products

A. Instatrak 3500 Plus System

Lot or serial number

N/A

Model or catalog number

1006236-001

Companies
Manufacturer

GE OEC Medical Systems Inc.