Health product recall

Insight and Ultra-Speed (2014-07-02)

Starting date:
July 2, 2014
Posting date:
July 10, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40407

Recalled Products

A) Insight

B) Ultra-Speed

Reason

Carestream has identified an incorrect orientation of the indicator dot on dental intraoral film packet labeling of some lots of insight, item 811 0785 (size 0 packet), ib-01, and some lots of ultra-speed, item 834 8658 (size 0 packet), df-48. The indicator dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the film. This issue may cause some confusion for the dental practitioner if the processed film is placed on the view box by referencing the film's raised dot. The result is the image orientation will be rotated 180 degrees. The remaining content on the label is correct. Image quality is not affected by this issue.

Affected products

A) Insight

Lot or serial number

538-013-03

539-012-01

545-012-01

546-012-01

Model or catalog number

811 0785

Companies
Manufacturer

Carestream Health Inc.

150 Verona Street

Rochester

14608

New York

UNITED STATES



B) Ultra-Speed

Lot or serial number

333-012-01

333-012-02

333-012-03

343-015-01

Model or catalog number

834 8658

Companies
Manufacturer

Carestream Health Inc.

150 Verona Street

Rochester

14608

New York

UNITED STATES