Health product recall

INNOVA E4 WASHER-DISINFECTOR (2016-03-16)

Starting date:
March 16, 2016
Posting date:
April 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57744

Affected Products

INNOVA E4 WASHER-DISINFECTOR

Reason

A hospital contacted SciCan on March 10, 2016 to report difficulties with one of their E4-WD Washer-Disinfectors. The hospital's biomedical technician reported that he identified the upper water flap of one channel being stuck in the closed position. According to the hospital's biomedical technician, a screw securing the water flap was suspected to be responsible for this issue. That specific screw was apparently loose and deflected the stainless steel insert spacer from its normal position causing it to miss the water flap which consequently remained closed. As the water flap was stuck in the closed position, it did not allow the proper circulation of water and chemicals in the upper scope. The scopes located in that specific location were not reprocessed correctly. The biomedical technician has confirmed that this water flap was opening correctly until March 9th, 2016.

Affected products

INNOVA E4 WASHER-DISINFECTOR

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

E4-WD

Companies
Manufacturer

BHT HYGIENETECHNIK GMBH

MESSERSCHMITTSTRASSE 11

GERSTHOFEN

86368

GERMANY