INNOVA E4 WASHER-DISINFECTOR (2016-03-16)
- Starting date:
- March 16, 2016
- Posting date:
- April 4, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57744
Affected Products
INNOVA E4 WASHER-DISINFECTOR
Reason
A hospital contacted SciCan on March 10, 2016 to report difficulties with one of their E4-WD Washer-Disinfectors. The hospital's biomedical technician reported that he identified the upper water flap of one channel being stuck in the closed position. According to the hospital's biomedical technician, a screw securing the water flap was suspected to be responsible for this issue. That specific screw was apparently loose and deflected the stainless steel insert spacer from its normal position causing it to miss the water flap which consequently remained closed. As the water flap was stuck in the closed position, it did not allow the proper circulation of water and chemicals in the upper scope. The scopes located in that specific location were not reprocessed correctly. The biomedical technician has confirmed that this water flap was opening correctly until March 9th, 2016.
Affected products
INNOVA E4 WASHER-DISINFECTOR
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
E4-WD
Companies
- Manufacturer
-
BHT HYGIENETECHNIK GMBH
MESSERSCHMITTSTRASSE 11
GERSTHOFEN
86368
GERMANY