Inhibin B Gen II ELISA
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Inhibin B Gen II ELISA |
240918D |
A81303 |
Issue
Beckman Coulter has determined that the affected Inhibin B Gen II ELISA reagent lots may provide elevated results. This increase is not detectable by kit controls. Their investigational data confirmed an average increase of about 15% with the maximum reported increase of an individual sample of +44%.
Recall Start Date: February 17, 2025
Additional information
Details
Immunotech S.R.O.
Radiova 1122/1, Praha 10, Czech Republic, 10200
Get notified