This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Ingenuity CT System (2015-02-19)

Starting date:
February 19, 2015
Posting date:
March 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52587

Recalled products

  1. Ingenuity CT System

Reason

Philips has become aware of the following issues:

Slab viewer longitudinal location issue - it is possible for the system to invert the sign (positive/negative) of the longitudinal position value displayed on an axial scan image. The following key events must take place for harm to occur:

  •  The patient is placed in a "feet-first" orientation.
  •  An interventional CT procedure is scheduled.
  • Contrary to typical clinical practice, a multi-planar reconstruction (MPR) is used for planning the interventional procedure (this issue only occurs in images produced using MPR).
  •  A check scan is not performed (normal clinical practice indicates the use of a check scan before beginning an interventional procedure).
  •  The needle tip placement is not confirmed as directed (this practice would prevent more than a minimal ingress of an intervention at an undesired location).
  •  Fast get ready issue - fast get ready incorrectly enabled during service procedures.
  •  Contrast annotation issue - contrast annotation incorrectly missing for some images when manual contrast was administered.
  •  Dot artifact issue - dot artifact present, intermittently, after startup.
  •  Heart rate update issue - heart rate measured value incorrectly remains constant for some heart rate change conditions.
  •  Doseright issue - doseright incorrectly enabled when the exam card specified disabling doseright.
  • Tube heat issue - tube heat predictor fails to warn of overheat condition for long scans.

Affected products

A. Ingenuity CT System

Lot or serial number

All lots

Model or catalog number
  • 728323
  • 728326
Companies
Manufacturer

Philips Medical Systems (Cleveland)

Inc.

595 Miner Road

UNITED STATES