Ingenuity CT System (2015-02-19)
- Starting date:
- February 19, 2015
- Posting date:
- March 20, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52587
Recalled products
- Ingenuity CT System
Reason
Philips has become aware of the following issues:
Slab viewer longitudinal location issue - it is possible for the system to invert the sign (positive/negative) of the longitudinal position value displayed on an axial scan image. The following key events must take place for harm to occur:
- The patient is placed in a "feet-first" orientation.
- An interventional CT procedure is scheduled.
- Contrary to typical clinical practice, a multi-planar reconstruction (MPR) is used for planning the interventional procedure (this issue only occurs in images produced using MPR).
- A check scan is not performed (normal clinical practice indicates the use of a check scan before beginning an interventional procedure).
- The needle tip placement is not confirmed as directed (this practice would prevent more than a minimal ingress of an intervention at an undesired location).
- Fast get ready issue - fast get ready incorrectly enabled during service procedures.
- Contrast annotation issue - contrast annotation incorrectly missing for some images when manual contrast was administered.
- Dot artifact issue - dot artifact present, intermittently, after startup.
- Heart rate update issue - heart rate measured value incorrectly remains constant for some heart rate change conditions.
- Doseright issue - doseright incorrectly enabled when the exam card specified disabling doseright.
- Tube heat issue - tube heat predictor fails to warn of overheat condition for long scans.
Affected products
A. Ingenuity CT System
Lot or serial number
All lots
Model or catalog number
- 728323
- 728326
Companies
- Manufacturer
-
Philips Medical Systems (Cleveland)
Inc.
595 Miner Road
UNITED STATES