Health product recall

Ingenio Family DREL Pacemakers and CRT- Ps (2021-06-03)

Starting date:
June 3, 2021
Posting date:
June 30, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-75923

Last updated: 2021-06-30

Affected Products

A. Advantio Pacemaker
B. Ingenio Pacemaker
C. Invive Cardiac Resynchronization Therapy Pacemaker
D. Vitalio Pacemaker
E. Inliven Cardiac Resynchronization therapy Pacemaker

Reason

L’impédance de la batterie EL augmente avec le temps, en fonction de la durée de l’implant et de la consommation d’énergie. Cette augmentation de l’impédance de la batterie peut entraîner des baisses de tension transitoires du dispositif pendant les périodes de forte consommation d’énergie associées à la communication télémétrique au moyen d’un programmateur ou d’un communicateur LATITUDE. L’appareil doit être remplacé.

Affected products

A. Advantio Pacemaker

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

K084

Companies
Manufacturer
Boston Scientific Corporation
4100 Hamline Avenue North
St. Paul
55112-5798
Minnesota
UNITED STATES

B. Ingenio Pacemaker

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

K184

Companies
Manufacturer
Boston Scientific Corporation
4100 Hamline Avenue North
St. Paul
55112-5798
Minnesota
UNITED STATES

C. Invive Cardiac Resynchronization Therapy Pacemaker

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

V182
V183

Companies
Manufacturer
Boston Scientific Corporation
4100 Hamline Avenue North
St. Paul
55112-5798
Minnesota
UNITED STATES

D. Vitalio Pacemaker

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

K284

Companies
Manufacturer
Boston Scientific Corporation
4100 Hamline Avenue North
St. Paul
55112-5798
Minnesota
UNITED STATES

E. Inliven Cardiac Resynchronization therapy Pacemaker

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

V284
V285

Companies
Manufacturer
Boston Scientific Corporation
4100 Hamline Avenue North
St. Paul
55112-5798
Minnesota
UNITED STATES