Ingenia 3.0T
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Ingenia 3.0T |
s/n: 34055 |
782103 |
Issue
Philips has identified an issue in the MR system patient support where the integrated radio frequency (IRF) carrier assembly screws may come loose and protrude beyond the surface. This could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop movement. The IRF carrier, a component of the patient support within the MR system, supports the tabletop where the patient lies. As the patient moves into the MRI tunnel, the tabletop, supported by the IRF carrier, forms a container for positioning the posterior coil.
Recall start date: September 4, 2024
Additional information
Details
| Philips Medical Systems Nederland B.V. |
| Veenpluis 6, Best, Netherlands, 5684 PC |
Get notified
Receive emails about new and updated recall and safety alerts.