Ingenia 3.0T
Last updated
Summary
Product
Ingenia 3.0T
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Ingenia 3.0T | s/n: 34055 | 782103 |
Issue
Philips has identified an issue in the MR system patient support where the integrated radio frequency (IRF) carrier assembly screws may come loose and protrude beyond the surface. This could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop movement. The IRF carrier, a component of the patient support within the MR system, supports the tabletop where the patient lies. As the patient moves into the MRI tunnel, the tabletop, supported by the IRF carrier, forms a container for positioning the posterior coil.
Recall start date: September 4, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
| Philips Medical Systems Nederland B.V. |
| Veenpluis 6, Best, Netherlands, 5684 PC |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76041