Health product recall

Infinity® Central Station

Dernière mise à jour

Summary

Produit
Infinity® Central Station
Problème
Medical devices - Performance issue
Ce qu’il faut faire

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Infinity® Centralstation Wide

All lots.

MS26800

Infinity® Central Station

All lots.

MS 18500
MS26800

Infinity® M300+

All lots.

MS33648
MS33659
MS33950

Infinity® M300

All lots.

MS 25755
MS18501
MS 26076

Problème

During our global market surveillance activities, we have become aware of cases in which the software on the Infinity® Central Station drops peaks on narrow waveforms causing the Infinity® M300 to fail the requirements of the standard IEC 60601-2-27.

Recall start date: March 9, 2023

Additional information

Details
Original published date: 2023-03-22
Type d’avis ou de rappel
Health product recall
Catégorie
Health product - Medical device - Cardiovascular
Health product - Medical device - General hospital and personal use
Entreprises

Draeger Medical Systems, Inc.

3135 Quarry Road, Telford, Pennsylvania, United States, 18969

Published by
Health Canada
Audience
Healthcare
Classe de rappel
Type I
Identification number
RA-73365