Infinity M500 Docking Station
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Infinity M500 Docking Station |
More than 10 numbers, contact manufacturer. |
MS20407 |
Issue
Draeger Medical Systems, Inc. has become aware during testing in conjunction with global market surveillance activities that the Infinity M500 Docking Station is not in full compliance to type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards.
The only ports that could be affected are: Hemo port (gray color), SpO2 port (blue color) and Temp port (white color) on M540. The specific accessories which can be connected to these ports and may be affected are the Dual Hemo MCable, SpO2 MCables, and temperature probes, respectively.
Recall Start Date: April 15, 2024
Additional information
Details
Draeger Medical Systems, Inc.
3135 Quarry Road, Telford, Pennsylvania, United States, 18969