Health product recall

Infinity Delta Patient Monitor (2017-03-31)

Starting date:
March 31, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63496

Affected Products

A. Infinity Delta Patient Monitor

B. Infinity Delta XL Patient Monitor

Reason

Draeger was informed about a case in which the affected units running VF10.0 software did not generate a set low O2 alarm although the O2 level that was measured by the Draeger Scio gas measurement module was below the adjusted alarm limit. Draeger analysis revealed a failure within the VF10.0 software. Earlier software versions are not affected.

Affected products

A. Infinity Delta Patient Monitor

Lot or serial number

5396182958

6008355470

6008535267

6008537461

6008537862

6008539566

Model or catalog number

MS18597

Companies
Manufacturer

Draeger Medical Systems, Inc.

3135 Quarry Road

Telford

18969

Pennsylvania

UNITED STATES