Infinity Central Station (2020-01-07)
- Starting date:
- January 7, 2020
- Posting date:
- February 7, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72257
Last updated: 2020-02-07
Affected Products
Infinity Central Station
Reason
Draeger received 16 complaints reporting a gray screen that intermittently appears when selecting the full disclosure feature utilizing the trending data function on the Infinity Central Sttation (ICS); patient monitoring continues as the user is able to exit the full disclosure feature, to restore the full disclosure functionality, a controlled reboot must be performed to ensure all patients are properly monitored during the reboot. Internal investigation confirmed a software defect manifested in version VG2.1.2.
Affected products
Infinity Central Station
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
MS 18500
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES