Health product recall

In2Flow Nasal Cannulas

Last updated

Summary

Product
In2Flow Nasal Cannulas
Issue
Medical devices - Performance issue
What to do

Customers provided instructions on how to respond when this becomes a problem. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

In2Flow Nasal Cannulas

1900059952
1900062390
1900059288
1900086945

10076606
10076605
10076604

Issue

Failure description: several complaints were received where the plug (the one at the opposite side of the tubing) became detached from the In2Flow nasal cannula, causing leakage due to unintentional disconnection of that part. This may result in degraded high flow support and oxygen delivery, which may lead to patient O2 desaturation.

Recall Start Date: August 31, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Hamilton Medical AG

Via Crusch 8, Bonaduz, Switzerland, 7402

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74337

Get notified

Receive notifications for new and updated recalls and alerts by category.