IMRIS iMRX System (2019-11-26)
- Starting date:
- November 26, 2019
- Posting date:
- January 6, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71997
Last updated: 2020-01-07
Affected Products
IMRIS iMRX System
Reason
It was found that the generator A100 may fail above a specific value due to a tolerance problem in the power supply of the generator control. The result is that the required voltage for the x-ray tube would not be supplied any longer. The issue occurs sporadically. However, it may occur during an on-going procedure which may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system. No reports of serious injury or death have been received by the company to date.
Affected products
IMRIS iMRX System
Lot or serial number
153374
Model or catalog number
10094141
Companies
- Manufacturer
-
Deerfield Imaging, Inc.
5101 Shady Oak Road
Minnetonka
55343
Minnesota
UNITED STATES