This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Important Safety Information on the Interactions between Active Implantable Medical Devices and Systems and other Medical Devices - Revision - Notice to Hospitals
- Starting date:
- October 31, 2005
- Posting date:
- October 31, 2005
- Type of communication:
- Notice to Hospitals
- Subcategory:
- Medical Device
- Source of recall:
- Health Canada
- Audience:
- Hospitals
- Identification number:
- RA-170001075
Notice about Health Canada advisories
Notice to Hospitals - Health Canada Issued Important Safety Information on Active Implantable Medical Devices and Systems
An updated Notice to Hospitals has been posted on Health Canada web site on December 20, 2005.
October 31, 2005
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Physiotherapy, Occupational Therapy, Radiology, Rehabilitative Medicine, Surgery, Neurosurgery, Orthopedic Surgery, Biomedical Engineering, Emergency Medicine, Pediatrics, Geriatrics, Internal Medicine, Nursing, Dentistry, Intensive Care and other involved professional staff and post this NOTICE in your institution.
Subject: Interactions between active implantable medical devices and systems and other medical devices
This is a revision of the information in the Health Canada Notice to Hospitals dated October 5, 2005.
Patients with active implantable medical devices and systems (AIMDS) are at risk of injury from interactions between the AIMDS and other medical devices used for diagnostic or therapeutic purposes. Such interactions could damage some AIMDS resulting in loss of therapy from these units, and could cause tissue damage resulting in severe injury or death. Patients may be at risk even if the leads are not connected to an implantable device or even if the device is off.
On April 2, 2003, Heath Canada issued a Notice to Hospitals entitled "Important safety information regarding DIATHERMY THERAPY in patients with implanted leads and implanted systems with leads".Footnote 1 It warned about the risks of tissue overheating, serious injury and death in this group of patients when subject to shortwave (radiofrequency) or microwave diathermy therapy. Since then, Health Canada has become aware of further international reports of serious injury resulting from interactions between AIMDS and other medical devices.Footnote 2
In 2003, the French Health Products Safety Agency, Agence française de sécurité sanitaire des produits de santé (AFSSAPS), assembled a working group to analyse the recommendations of AIMDS manufacturers and make recommendations to health care professionals who perform certain diagnostic and therapeutic procedures on patients with AIMDS. The recommendations of the group were released by AFSSAPS in February 2005.Footnote 3 Health Canada has reviewed these recommendations and is presenting them in an adapted form as follows:
| Implanted Cardiac Pacemaker | Implanted Cardiac Defibrillator | Implanted Neurostimulator | |
|---|---|---|---|
|
Table 1 footnotes
|
|||
| Shortwave and Microwave Diathermy | Absolute contraindication | Absolute contraindication | Absolute contraindication |
| Ultrasound Diathermy | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Absolute contraindication |
| Electromagnetic Stimulation Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Absolute contraindication |
| MRI | Absolute contraindication | Absolute contraindication | Absolute contraindication |
| Electrosurgery Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| External Defibrillator | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| Radiotherapy Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| Lithotripsy or Ultrasound Therapy Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| Fluoroscopy or other X-ray Devices | No contraindication | No contraindication | No contraindication |
| Echography | No contraindication | No contraindication | No contraindication |
The risk of interactions between AIMDS and other medical devices used for diagnostic or therapeutic purposes depends on many parameters. Due to the diversity and complexity of the devices in question, Health Canada wishes to pass along these general guidelines but cannot give all-encompassing recommendations that cover all possible clinical scenarios. Health Canada continues to monitor this evolving safety issue and will release more information as it becomes available.
Health Canada depends on health care professionals to report adverse incidents related to medical devices. Any serious and/or unexpected adverse incident related to medical devices should be reported to Health Canada at the following address:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
For other inquiries, please refer to contact information:
Marketed Health Products Directorate (MHPD)
Email: MHPD_DPSC@hc-sc.gc.ca
Tel: (613) 954-6522
Fax: (613) 952-7738
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
Footnotes
- Footnote 1
-
Important safety information regarding DIATHERMY THERAPY in patients with implanted leads and implanted systems with leads.
Notice To Hospitals. Health Products and Food Branch, Health Canada. April 2, 2003. - Footnote 2
-
FDA Public Health Notification: MRI-Caused Injuries in Patients with Implanted Neurological Stimulators.
Center for Devices and Radiological Health, Food and Drug Administration. May 10, 2005. - Footnote 3
-
Interactions entre dispositifs médicaux implantables actifs et dispositifs médicaux.
[NJS_FILE:3311bdad-28b4-40a2-8077-a66dbc3cbc0f:98fdee3d-f9e9-430d-9588-8eb54923e2cb].