Important Safety Information - BHM/Medi-Man Combi Sling - Possible Stitching Failure - Notice to Hospitals
- Starting date:
- April 3, 2009
- Posting date:
- May 14, 2009
- Type of communication:
- Notice to Hospitals
- Subcategory:
- Medical Device
- Source of recall:
- Health Canada
- Audience:
- Healthcare Professionals
- Identification number:
- RA-170001983
This is duplicated text of a letter from BHM Medical inc.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
Notice to Hospitals - Health Canada Endorsed Important Safety Information on BHM/Medi-Man Combi Sling
April 3, 2009
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Surgery; Emergency Medecine, Pediatrics, Geriatrics, Internal Medicine, Nursing, Intensive Care and/or other Departments as required, and other involved professional staff and post this Notice in your institution.
Subject: BHM/Medi-Man Combi Sling – Possible Stitching Failure
Product Family: Sling
Document Number: FSN10/2009/ca
Distribution: All end users
Scope
| Part number | Description |
|---|---|
| 626002 | COMBI SLING |
| 626002M | COMBI SLING MESH DELUXE |
| 626002C | COMBI SLING DELUXE CHILD |
| 626002C-M | COMBI SLING MESH DELUXE CHILD |
| 626003 | COMBI SLING DELUXE OVERSIZE |
| 626003M | COMBI SLING MESH DELUXE OVERSIZE |
| 626003X | COMBI SLING OVERSIZE HEAVY DUTY |
Combi slings can be identified by looking for the following characteristics:
Characteristics
- Yellow, green and white shoulder straps
- Handle in the back
- Green strap in the back
- White main color
- Orange, black and white leg straps
These slings were labeled under the BHM or Medi-Man brand.
Background
This notice relates to premature failure of the stitching at the junction of the shoulder strap and the sling body which can be noticed during routine inspection. The failure of the stitching is not sudden but gradual; therefore it is highly unlikely that a sling that has been inspected and considered suitable for use prior to a transfer would fail during that transfer.
Health Hazard Analysis
An undetected stitching failure can ultimately lead to the detachment of the shoulder strap resulting in a possible patient fall. This may lead to serious personal injuries or worse to both patient and/or caregiver.
Actions to Take
- Make sure you carefully read fully and understand this field safety notice.
- Make sure this notice is distributed to all those who need to be aware within your organization or to any organization where the potentially affected slings have been transferred.
-
Inspect and assess your slings suitability for use as per the Slings Maintenance Instruction (001.03685) originally provided with your sling. Look closely for thread breakage at the junction of the shoulder strap and the sling body as shown in the Figure 2. If your sling is showing this kind of breakage, remove your sling from use and discard it immediately. If the inspection reveals no deterioration of the stitching, the sling is safe to use and can be returned to use.
Figure 2: Example of partial thread breakage
- Follow the General Care Guidelines section of the Slings Maintenance Instructions, namely the directives "It is imperative that a patient transfer sling be inspected prior to each use."
Next Step
Actions have already been initiated for the redesign of the Combi sling. The improved Combi Sling is expected to be available in May 2009.
You will be contacted for the replacement of your eligible old slings. Slings which have been in use for up to two (2) years will be replaced at no charge.
If you have any further questions or special needs, please contact our service department at:
BHM Medical Inc.
2001 Tanguay, Magog, QC, Canada J1X 5Y5
Telephone: 819-868-0441 Toll free: 1-800-868-0441
Fax: 819-868-2249
E-mail: service@bhm-medical.com
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of serious specific adverse incident or other serious or unexpected adverse incidents in patients using BHM/Medi-Man Combi Sling should be reported to BHM Medical Inc. or Health Canada at the following address:
BHM Medical Inc.
2001 Tanguay, Magog, QC, Canada J1X 5Y5
Telephone: 819-868-0441 Toll free: 1-800-868-0441
Fax: 819-868-2249
E-mail: service@bhm-medical.com
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: mhpd_dpsc@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
Long Descriptions
- Long Description 1
-
Combi slings characteristics
The picture shows the sling, with arrows describing each part of the sling in detail and pointing to those parts. There is an actual picture of the sling on the right, and a diagram picture of the same sling on the left for better detail.
- Long Description 2
-
Premature failure of the stitching
A picture of the sling, with a caption on the right showing the zoomed image of the stitching junction where the failure may occur.
- Long Description 3
-
Premature failure of the stitching
A larger zoomed image of the part of the sling showing example of stitching failure.
Images
Select thumbnail to enlarge - opens in a new window