Impella® Controller
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Impella® Controller |
More than 10 numbers, contact manufacturer. |
0042-0040-CA |
Impella® Controller |
More than 10 numbers, contact manufacturer. |
0042-0000-CA |
Issue
Abiomed has identified an Automated Impella Controller™ (AIC) issue that may impair the detection of an Impella® pump when connected to an AIC. The pump detection issue may occur with any of the Abiomed Impella® pumps and may occur during console-to-console transfer or case start. There is no visual alarm displayed on the AIC screen to indicate the detection issue in these situations.
Recall start date: July 7, 2025
Additional information
Details
Abiomed Inc.
22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923