Impella CP With SmartAssist and Impella 5.5 With SmartAssist
Brand(s)
Last updated
Summary
Product
Impella CP With SmartAssist and Impella 5.5 With SmartAssist
Issue
Medical devices - Device compatibility
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Impella CP With SmartAssist | All lots. | 0048-0044 |
| Impella 5.5 With SmartAssist | All lots. | 0550-0004 |
Issue
Abiomed has determined there is a risk of optical sensor damage in Impella products when used concurrently and in close proximity with shockwave coronary intra-vascular lithotripsy catheter.
Recall Start Date: September 3, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Abiomed Inc.
22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-76002
Get notified
Receive emails about new and updated recall and safety alerts.