ImmunoCAP Allergen c6, Amoxicilloyl and c5, Ampicilloyl (2017-12-28)
- Starting date:
- December 28, 2017
- Posting date:
- January 12, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65684
Affected products
- ImmunoCAP Allergen c6, Amoxicilloyl
- ImmunoCAP Allergen c5, Ampicilloyl
Reason
Somagen Diagnostics along with Thermofisher Scientific would like to notify you of an urgent recall and backorder of ImmunoCAP Allergen c5, Ampicilloyl product code 14-4450-01 and ImmunoCAP Allergen c6, Amoxicilloyl product code 14-4451-01. An internal monitoring stability study has indicated that stability specifications have not been met and were in fact half the intended expiry date of 12 months. There is a possibility that samples with a low positive concentration may be incorrectly reported as negative. Any reported test results below the assay Limit of Qualification (LOQ) should be disregarded as these results may be invalid. No complaints or injuries to patients have been reported.
Affected products
A. ImmunoCAP Allergen c6, Amoxicilloyl
Lot or serial number
- 991D7
- 991D8
- 991D9
- 991DA
- 991DB
- 991DC
Model or catalog number
14-4451-01
Companies
- Manufacturer
-
PHADIA AB
RAPSGATAN 7, P.O. BOX 6460
UPPSALA
75137
SWEDEN
B. ImmunoCAP Allergen c5, Ampicilloyl
Lot or serial number
- 990DS
- 990DT
- 990DU
- 990DV
- 990DW
- 990DX
- 990DY
- 990DZ
Model or catalog number
14-4450-01
Companies
- Manufacturer
-
PHADIA AB
RAPSGATAN 7, P.O. BOX 6460
UPPSALA
75137
SWEDEN