Immage Immunochemistry Systems Immunoglobulin G Reagent (IgG)

Starting date:
February 6, 2017
Posting date:
February 27, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62392

Reason

Immage IgG Lot M610132A parameter card barcode does not include the "a" suffix encoded in the barcode.

The reagent cartridge barcode needed to pair with the parameter card includes the "a" suffix.

Lot number mismatch causes a "parameters not found" error upon reagent cartridge barcode read.

Input of parameters and reagent load is by barcode only. System is not capable of manual entry.

Affected products

Immage Immunochemistry Systems Immunoglobulin G Reagent (IgG)

Lot or serial number

M610132A

Model or catalog number

446400

Companies

Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821