IMMAGE Immunochemistry System Total Immunoglobulin E (IGE) Reagent and Cal (2019-09-04)
- Starting date:
- September 4, 2019
- Posting date:
- October 3, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71129
Last updated:
Affected Products
IMMAGE Immunochemistry System Total Immunoglobulin E (IGE) Reagent and Cal
Reason
Patient samples with IGE concentrations in the assay extended analytical range (>500- 30,000 IU/ml) may generate false low results that recover within the initial undiluted (neat) range of 5-500 IU/ml. Diagnosis and treatment may be delayed if results are not evaluated in conjunction with patient symptoms or other clinical tests.
Affected products
IMMAGE Immunochemistry System Total Immunoglobulin E (IGE) Reagent and Cal
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
474620