Health product recall

IMMAGE Immunochemistry System IgM Immunoglobulin M Reagent (2019-04-21)

Starting date:
April 21, 2019
Posting date:
June 28, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70327


Last updated:

Affected Products

IMMAGE Immunochemistry System IgM Immunoglobulin M Reagent

Reason

Patient samples tested with IgM reagent lots M803185 and M806164 may generate E60 (no reaction, antigen excess failed) or E66 (unstable reaction) error messages on very low concentration samples.

Affected products

IMMAGE Immunochemistry System IgM Immunoglobulin M Reagent

Lot or serial number

M803185

M806164

Model or catalog number

447610

Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES