Issue
Following customers' feedback, we have been able to confirm that the results obtained with this Ih-Qc 3 batch may not be as expected. Different issues were observed which may cause Qc failed results:
a. Positive reactions may be observed in Direct Antiglobulin Tests (DAT)
b. Weaker or in worse case false negative reactions may occur with Rh:4 cells in Indirect Antiglobulin Test (IAT) and in 2-stage papain technique
c. When testing the impacted batch of Ih-Qc3 with Id-cards containing Anti-C, a double population with Anti-C (Rh:4) can be observed
Recall start date: Apr 25, 2022
Additional information
Details
Diamed Gmbh
Pra Rond 23, Cressier Fr
Switzerland, 1785