Health product recall

Idylla Instrument

Brand(s)
Last updated

Summary

Product
Idylla Instrument
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Idylla Instrument

SER2149
SER3468
SER2361
SER3460
SER3302
SER3321
SER3322
SER2501

P0010

Issue

Biocartis received a complaint on several Idylla MSI test false positive results generated on a single instrument. As part of the investigation, Biocartis conducted a review of the Idylla MSI test performance in the field. A review of all connected runs from 2019 to date identified suspected false positive Idylla MSI test runs obtained by customers in Germany, Italy, Poland, Estonia, Cyprus, and Thailand.

Recall start date: December 9, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Biocartis Nv
Generaal De Wittelaan, Mechelen, Belgium, 2800
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76770