This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

IDS-ISYS-Wash system (2014-01-24)

Starting date:
January 24, 2014
Posting date:
March 20, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38447

Recalled Products

IDS-ISYS-Wash system

Reason

The ID’s continual quality monitoring program has identified that a very small number of wash solution containers have been manufactured without adding the concentrated solution. The potential hazard associated with the use of the device is an increase in the imprecision and a bias relative to the true concentration of samples measured using the IDS-ISYS assays on board the IDS-ISYS analyser.

Affected products

IDS-ISYS-Wash system

Lot or serial number

IS-CW100

Model or catalog number

75496

76208

Companies
Manufacturer

Immunodiagnostic Systems Limited

10 Didcot Way, Boldon Business Park,

Boldon, Tyne and Wear,

UNITED KINGDOM