Health product recall

iChem 10 SG Urine Chemistry Strips (2014-12-09)

Starting date:
December 9, 2014
Posting date:
December 29, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43039

Recalled Products   

iChem 10 SG Urine Chemistry Strips 

Reason

The identified lots of iChem 10 SG Urine Chemistry Strips (only lot 510033A was shipped to Canada) may have leukocyte pads that are more sensitive at the first reporting range of 1+, corresponding to 25 white blood cells per microliter, leading to false positive leukocyte results for samples that are truly negative.

Affected products

iChem 10 SG Urine Chemistry Strips 

Lot or serial number
  • 510033A
  • 510034A
  • 510035A
  • 510035B
Model or catalog number

800-7005

Companies
Manufacturer

Iris Diagnostics, a Division of Iris International Inc.

9172 Eton Avenue,

Chatsworth

91311

California

UNITED STATES