Health product recall

IACS MONITORING SOLUTION WITH C500, C700 & INFINITY M540 (2020-12-30)

Starting date:
December 30, 2020
Posting date:
January 22, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74821



Last updated:
2021-01-22

Affected Products 

  1. IACS MONITORING SOLUTION WITH C500
  2. IACS MONITORING SOLUTION WITH C700
  3. INFINITY M540

Reason

The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.

Affected products

A. IACS MONITORING SOLUTION WITH C500

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

MS25510

Companies
Manufacturer

Draeger Medical Systems, Inc.

3135 Quarry Road

Telford

18969

Pennsylvania

UNITED STATES


B. IACS MONITORING SOLUTION WITH C700

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

MS25520

Companies
Manufacturer

Draeger Medical Systems, Inc.

3135 Quarry Road

Telford

18969

Pennsylvania

UNITED STATES


C. INFINITY M540

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

MS20401

Companies
Manufacturer

Draeger Medical Systems, Inc.

3135 Quarry Road

Telford

18969

Pennsylvania

UNITED STATES