Health product recall

IACS Monitoring Solution with C500 and C700, and Infinity M540 (2018-02-28)

Starting date:
February 28, 2018
Posting date:
May 25, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66880

Affected products

  1. IACS Monitoring Solution with C500
  2. IACS Monitoring Solution with C700
  3. Infinity M540

Reason

When the Infinity M540 patient monitor is placed in standby mode and the user simultaneously connects the Infinity MCable - Dual Hemo or Infinity MPod – Quad Hemo to the Infinity M540 patient monitor and then proceeds to zero the invasive pressure channel, in rare cases the invasive pressure visual and audible alarms are not activated. Although the invasive pressure value displayed on the Infinity M540 patient monitor and on the infinity medical cockpit is accurate, the invasive pressure visual and audible alarms will stay disabled until the patient is discharged or the hemodynamic pod is disconnected and reconnected.

Affected products

A. IACS Monitoring Solution with C500

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

MS25510

Companies
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
UNITED STATES

B. IACS Monitoring Solution with C700

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

MS25520

Companies
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
UNITED STATES

C. Infinity M540

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

MS20401
MS22956
MS25643
MS26196
MS31818

Companies
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
UNITED STATES