IACS Monitoring Solution with C500 and C700 (2017-06-12)
- Starting date:
- June 12, 2017
- Posting date:
- July 7, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals
- Identification number:
- RA-63902
Affected Products
A. IACS Monitoring Solution with C500
B. IACS Monitoring Solution with C700
Reason
The Infinity Acute Care System Monitoring Solution Cockpits (IACS Monitoring Solution with C500 and IACS Monitoring Solution with C700) with Revision Index 6 or higher that contain 4GB RAM modules may induce a failure that can lead to high memory usage thereby delaying touchscreen operations, and may also lead to not annunciating audio or visual alarms on the cockpit and infinity central station. The M540 will still issue visual alarms.
Affected products
A. IACS Monitoring Solution with C500
Lot or serial number
376587013
402073021
402073030
402073067
Model or catalog number
MS25510
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES
B. IACS Monitoring Solution with C700
Lot or serial number
350603074
352066008
354380009
Model or catalog number
MS25520
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES