Health product recall

HVAD Implant Kit

Brand(s)
Last updated

Summary

Product
HVAD Implant Kit
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
HVAD Implant Kit More than 10 numbers, contact manufacturer. 1101 1102 1103 1104 1205 1104CA-CLIN

Issue

Pump thrombosis caused by weld defect that allowed moisture into the center post and corroded the magnets that keep the impeller rotating concentrically.

Recall start date: April 26, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Heartware Inc.

14400 Nw 60th Avenue, Miami Lakes, Florida, United States, 33014

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64201

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe