Health product recall

HLS Set Advanced Bioline Coating

Last updated

Summary

Product
HLS Set Advanced Bioline Coating
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

HLS Set Advanced Bioline Coating

3000183179
3000183910
3000183908
3000183172
3000183191
3000184681

70105-2794 BEQ-HLS 7050-USA
70105.2797 BEQ-HLS 5050-USA

Issue

The former packaging design of HLS set contains accessories packed in medical paper as opposed to the new introduced packaging design in Tyvek. Based on design verification tests it showed that medical paper is a weaker technical solution than Tyvek, carrying a potential risk of breaching and thus compromising the product sterility. Therefore it was decided to remove all HLS sets with former packaging design from the market. New packaging was introduced in September 2021. 

Recall start date: July 18, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Maquet Cardiopulmonary GmbH

Kehler Strasse 31, Rastatt, Germany, 76437

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64395

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe