Health product recall

HITACHI HICO Creatinine Jaffe Method, CREJ2 cobas (c 111, c 701/702, c 311/501/502), OCBAS INTEGRA 400 plus (2019-01-16)

Starting date:
January 16, 2019
Posting date:
March 1, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69172

Affected Products

  1. HITACHI HICO Creatinine Jaffe Method, Rate-Blanked & Compensated (CREA)
  2. CREJ2 cobas c 111
  3. CREJ2, cobas c 701/702
  4. CREJ2 cobas c 311/501/502/ COBAS INTEGRA 400 plus

Reason

Creatinine Jaffe was found to interfere with glucose, pyruvate and ascorbate. No interference was found for the creatinine enzymatic assays on all platforms.

Affected products

  1. HITACHI HICO Creatinine Jaffe Method, Rate-Blanked & Compensated (CREA)

Lot or serial number

21135801
22521601
23333001
24018201
24724601
25179801
28039501
29422901
33545001

Model or catalog number

11875418216
11875663216
11929941216
 

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY

B. CREJ2 cobas c 111

Lot or serial number

23421901
26739801
31788801
35247501

Model or catalog number

05401755190

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY

C. CREJ2, cobas c 701/702

Lot or serial number

24327001
25635101
27623601
29187701
31403501
32683201
34583301
35712601
37008301

Model or catalog number

06407137190
 

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY

D. CREJ2 cobas c 311/501/502/ COBAS INTEGRA 400 plus

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

04810716190
 

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY