Product
Histocore Pegasus
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Histocore Pegasus

G0061 - G0701

14048858009

Issue

Leica Biosystems became aware, through post market surveillance, of an issue regarding poorly processed and/or damaged biopsy tissue specimens on the Histocore Pegasus / Histocore Pegasus Plus, resulting from incorrect assignment of the carryover value set for created or edited protocols. If the setup of carryover is lower than the actual carryover, it may potentially cause tissue damage (mainly under-processed tissue).

Recall start date: November 9, 2023