This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

HerpeSelect 1 and 2 Immunoblot IgG

Starting date:
February 24, 2017
Posting date:
March 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62782

Reason

Diasorin Molecular have recently received customer complaints regarding IB0900G Lot# 32574 and 32865 producing strong common ag bands with no bands for glycoprotein G1 and G2. As a consequence, the lack of G1 and G2 bands has led to an increase in equivocal results.

Affected products

HerpeSelect 1 and 2 Immunoblot IgG

Lot or serial number

32574

32865

Model or catalog number

IB0900G

Companies

Manufacturer

Diasorin Molecular LLC (also trading as Focus Diagnostics, Inc.)

11331 Valley View Street

Cypress

90630

California

UNITED STATES